Tel:
+49 (0)241 95 163 153
Fax:
+49 (0)241 95 163 155
E-Mail:
orders@anticorps-enligne.fr

CRYBA1 Kit ELISA

Kit ELISA Souris CRYBA1, test Colorimetric pour la quantification de Souris CRYBA1.
N° du produit ABIN5653388

Aperçu rapide pour CRYBA1 Kit ELISA (ABIN5653388)

Antigène

Tous les produits CRYBA1
CRYBA1 (Crystallin, beta A1 (CRYBA1))

Reactivité

Souris

Méthode de détection

Colorimetric

Type de méthode

Sandwich ELISA

Gamme de detection

0.312 ng/mL - 20 ng/mL

Application

ELISA

Type d'échantillon

Cell Lysate, Tissue Homogenate
  • Seuil minimal de détection

    0.312 ng/mL

    Analytical Method

    Quantitative

    Specificité

    This assay has high sensitivity and excellent specificity for detection of Crystallin Beta A1 (CRYbA1). No significant cross-reactivity or interference between Crystallin Beta A1 (CRYbA1) and analogues was observed.

    Sensibilité

    0.118 ng/mL
  • Commentaires

    The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5 % within the expiration date under appropriate storage condition. To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

    Durée du test

    3 h

    Plaque

    Pre-coated

    Protocole

    The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Crystallin Beta A1 (CRYbA1). Standards or samples are then added to the appropriate microtiter plate wells with a biotin-conjugated antibody specific to Crystallin Beta A1 (CRYbA1). Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Crystallin Beta A1 (CRYbA1), biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Crystallin Beta A1 (CRYbA1) in the samples is then determined by comparing the O.D. of the samples to the standard curve.

    Précision du teste

    Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Crystallin Beta A1 (CRYbA1) were tested 20 times on one plate, respectively
    Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Crystallin Beta A1 (CRYbA1) were tested on 3 different plates, 8 replicates in each plate. CV(%) = SD/meanX100
    Intra-Assay: CV<10%
    Inter-Assay: CV<12%

    Restrictions

    For Research Use only
  • Conseil sur la manipulation

    The Stop Solution is acidic. Do not allow to contact skin or eyes. Calibrators, controls and specimen samples should be assayed in duplicate. Once the procedure has been started, all steps should be completed without interruption.

    Stock

    4 °C,-20 °C

    Stockage commentaire

    -20°C. Bring all reagents to room temperature before beginning test. The kit may be stored at 4°C for immediate use within two days upon arrival. Reseal any unused strips with desiccant pack. Minimize freeze/thaw cycles.

    Date de péremption

    4-8 months
  • Antigène Tous les produits CRYBA1

    CRYBA1 (Crystallin, beta A1 (CRYBA1))

    Autre désignation

    Crystallin Beta A1

    Sujet

    Gene Name: Crystallin Beta A1

    Gene Aliases: CRYB1, Eye Lens Structural Protein

    UniProt

    P02525
Vous êtes ici: